beautypg.com

1about this document – K-Patents PR-23-SD Generation 1 User Manual

Page 3

background image

© Copyright K-Patents 2013. All rights reserved.

3

Safe-Drive™ Process Refractometer

PR-23-SD Generation 1

Best Practices

1

About This Document

This document is intended for individuals
installing, commissioning, operating, and/
or servicing the Safe-Drive™ Process
Refractometer PR-23-SD generation 1
model that has been manufactured in
years 2006–2013. The purpose of this
document is to provide a quick guide for
the abovementioned tasks in the form of
K-Patents recommended best practices.

This document in intended for PR-23-SD applications that have a steam wash system
(typical application for black liquor).

NOTE: These instructions are for quick reference only. For more thorough guidance, please
refer to K-Patents user manual and documentation.

2

Before Installation

2.1 Installation Location Checklist

Before the SD sensor system is installed, it is important to inspect the installation location
carefully for the following conditions:

● Install the SD sensor system onto either vertical or horizontal pipeline. The

recommended flow rate is 0.4 m/s – 2 m/s (1.5 ft/s – 6 ft/s).

● Mount the system at waist level. This is the natural and safe height for the system and

enables you to use the tools more ergonomically.

● Leave a 1 m (approximately 3 ft) space around the installation for operating around the

SD sensor system. For the same reason, the steam connections should be installed
on the sides or back of the sensor system.

● Recommended maximum ambient temperature of installation location is 45°C (120°F).

● Avoid locations that are blocked by other piping and/or equipment, or require additional

tools, such as a ladder, to access the sensor. The installation location must be level,
firm, and free of clutter to provide safe and easy access to the system.

This manual is related to the following products: