Please observe the instructions of the manufacturer of the CAD/CAM system for the selection of the 
model material.

Important for the fabrication of models

:

−  Check the radius of the incisal/occlusal edge on the prepared tooth (maxilla and mandible).
−  The prepared incisal/occlusal edge should be at least as thick as the diameter of the bur used in the 

cavity during the CAD/CAM process.

−  If the incisal/occlusal edge of the prepared die is thinner than the diameter of the bur, the edge has to 

be blocked out accordingly.

−  The preparation margins must be clearly defined so that they are clearly recorded during scanning.
−  Do not apply a die varnish or spacer.
−  In clinical cases with a very thin alveolar ridge, the lingual/palatal side of the ridge must be 

 supplemented with putty silicone or a similar material. This allows the dimensions of the bridge pontic 
to be correctly designed during the subsequent design of the restoration.

A stone model is used as a working model.

In clinical cases with a very thin alveolar ridge, the lingual/palatal side must be blocked out on the model.