A thin layer of Multilink Hybrid Abutment is directly applied from the mixing tip to the

bonding surface of the Ti base.

The ceramic structure is placed on the Ti base in such a way that the position markings

are aligned. The components are joined using even and light pressure and the relative

position of the components is checked (transition base/ceramic structure).

Excess in the screw channel is carefully removed, e.g. with a Microbrush or brush,

using rotary movements.

Glycerine gel (e.g. Liquid Strip) is applied on the cementation joint to prevent the

formation of an inhibition layer.

The luting composite auto-polymerizes within 7 min.

Important: The components must not be moved until auto-polymerization

is completed. The components must be immobilized during this time.

A thin layer of Multilink Hybrid Abutment is directly applied from the mixing tip

on the bonding surface of the ceramic structure.

Subsequently, the components are tightly pressed together for 5 s.

Important: Excess must not be removed before curing has started,

i.e. 2–3 minutes after mixing. The components are held in place with

light pressure in the process.

5 s

2-3 m

7 min