The following procedure is recommended:
– For easier handling, the ceramic structure can be positioned on the Ti base for glazing. For that purpose, secure Ti base

on a model analog.

– Mix the glazing material (IPS e.max Ceram Glaze Paste or Powder) with the IPS e.max Ceram Glaze and Stain Liquid

allround or longlife to the desired consistency.

– Apply an even layer of glazing material covering all areas that are to be glazed.
– If required, the fluorescence may be increased by applying a fluorescing glazing material (paste or powder).

Important:

The glazing material must neither reach the bonding surface to the Ti base nor the screw channel, as this may
compromise the accuracy of fit. Check the interface before firing and carefully remove any contamination. On the
abutment, do not apply any glaze to the bonding surface to the crown, as this might compromise the fit of the
crown.


The IPS e.max CAD Crystall./Glaze Spray is not recommended for glazing IPS e.max CAD Abutment
Solutions, as is requires very targeted application. The glazing material must neither reach the bonding surface
to the Ti base nor the screw channel, as this may compromise the accuracy of fit.

Conduct the Characterization/Glaze firing for IPS e.max Ceram on a honey-comb firing
tray using the stipulated firing parameters. Firing parameters see page 64

An even layer of glaze material is applied to the emergence profile of the hybrid abut-

ment. Care has to be taken that no glaze material enters the screw channel.

Care has to be taken that no glaze material is present on the interface of the hybrid

abutment and hybrid abutment crown prior to the firing cycle. The glaze material is

carefully removed, if necessary.

The glazing material is applied evenly on the outer surface of the hybrid abutment

crown. Care has to be taken that no glaze material enters the screw channel.

The Characterization/Glaze firing is conducted on a honey-comb firing tray with the

corresponding parameters.