Layer thicknesses of the ceramic components

Observing the geometry requirements of the IPS e.max CAD ceramic structures is the key to success for a durable restoration.
The more attention is given to the design, the better the final results and the clinical success will turn out to be.

The following basic guidelines have to be observed:

Wall thickness W

HAK

Width

hybrid abutment crown B

HAK

Width crown B

Crown

Hybrid abutment crown

Hybrid abutment

– The wall thickness of hybrid abutment crown

W

HAK

must be larger than 1.5 mm for the entire

circumference.

– The opening of the screw channel must not be

located in the area of contact points. If this is not
possible, a hybrid abutment with a separate crown
would be preferred.

– The width of the hybrid abutment crown B

HAK

is

limited to 6.0 mm from the axial height of contour
to the screw channel.

– The notes of the implant manufacturer must be

observed regarding the maximum height of the
hybrid abutment crown.

– The wall thickness W

HA

, must be at least 0.5 mm.

– The hybrid abutment should be designed in a

similar way as a prepared natural tooth:

– Circular epi-/supragingival shoulder with rounded

inner edges or a chamfer.

– In order for the crown to be cemented to the

hybrid abutment using a conventional/self-
adhesive cementation protocol, retentive surfaces
and a sufficient "preparation height" must be
observed.

– Create an emergence profile with a right angle at

the transition to the crown (see picture).

– The crown width B

Crown

is limited to 6.0 mm from

the axial height of contour to the screw channel of
the hybrid abutment.

– The notes of the implant manufacturer must be

observed regarding the maximum height of the
hybrid abutment and separate crown.

Wall thickness W

HA