System -39 – AMETEK SGI Series User Manual
Page 91
Sorensen SGI Series
Calibration and Verification
M550221-01 Rev AA
3-39
GPIB
AND
RS232
Note: For Standard SGI without Ethernet
The Remote Menu for GPIB option or RS232 option
(Home Page 3, F1) allows setting of the GPIB address
and the baud rate for the RS232 interface. Use the
NavPad to set the GPIB address from 1 to 30, and the
RS-232 baud rate from 2400 to 19200. Pressing F1 saves
the new values in non-volatile memory so that they will be
remembered after power down. Pressing F4 keeps the
selected values only until the unit is power down; then, at
next power up, the previously saved settings will be
restored.
R
EMOTE
O
PERATION
S
CREENS
If the unit is being controlled remotely via Ethernet, GPIB or RS-232 interface,
special screens may appear during operation. The following describes those
displays:
Remote Mode Primary display shows the actual values
for Volts and Watts. F4 jumps to Home Timeout
screen.
Remote Mode Power display shows the set points on
the first line, actual values for Voltage and Current on
the second line, and at the bottom, the actual value for
Watts with a bar graph of its percent of full-scale power.
F4 jumps to the Constant Power screen.
Remote Mode Sequence display shows what sequence
is running. F4 jumps to the local Sequence Run
screen.
3.8.12 System
The System submenu allows the user to display the total
system current when the unit is being paralleled with
other SG series supplies.
From Home Menu Page 3, press F2 to enter the System Menu. Determine the
Total System Current by adding the maximum current rating of all supplies in
parallel. Use NavPad or the KeyPad "0-9" to enter the appropriate value. F3
resets Total System Current to the supply default value; F4 keeps the
displayed value and jumps to Home Menu Page 1.