Sorensen SGI Series

Calibration and Verification

M550221-01 Rev AA

3-51

25

14 ISO TTL/CMOS

1

ISO RTN 2

Figure 3-21. Remote Output On/Off Using Isolated TTL/CMOS Source

3.13.3 Remote Shutdown(S/D)

A remote +12 VDC voltage can be connected externally between Pin J1-18
(S/D Fault) and Pin J1-24 (COM) to disable, i.e., shut down the output of the
power supply; see Figure 3-22. A low-level, or opening the +12 VDC signal,
will allow the unit to revert to normal operation.

Figure 3-22. Remote Shutdown Using DC Voltage Source