Initial start-up displays -13, Initial start-up displays – AMETEK SGI Series User Manual
Page 65
Sorensen SGI Series
Calibration and Verification
M550221-01 Rev AA
3-13
9. Enter the Default Programming menu by pressing either
the voltage, current, or overvoltage front panel keys:
10. Program the OVP setting as appropriate for the
application as shown above. If OVP will not be used,
then “OVP set” programming may be set at maximum, approximately
110% of the rated output voltage of the supply.
11. If OVP mode did not function as indicated above, verify the setup and
perform the check again. If the function continues to fail, contact the
factory for assistance.
3.3
Initial Start-Up Displays
Upon initial start-up of unit, the following screens are displayed in sequence,
after initial splash screens which show manufacturer's information:
L
AST
C
AL
D
ATE
Displays for seven seconds, showing date that the unit was last calibrated.
See SG Series Programming Manual, P/N M550129-03, for calibration
information and how to update "Last Cal Date."
O
UTPUT
-E
NABLED
W
ARNING
Indicates number of seconds left in a 10-second countdown, warning the user
that the output will be enabled at the end of the countdown unless the process
is aborted by pressing the F1 key on the front panel. Once aborted, the output
remains off until the user enables it with the Output On/Off key: refer to
Section 3-1. This warning appears only if the unit is both in Local mode and
was last set and saved to enable the output at its next start-up. See
Section 3.8.2 for changing and saving the Power On (PwrOn) Output settings.
Note:
Factory default setting for Output at start-up is ON, output enabled.
H
OME
M
ENU
,
P
AGE
1
Final screen to display upon initial start-up of the unit; it is from here that the
user may access all menu functions. See Menu Map, Section 3.7, for Home