SOR Calibration Manager Software User Manual
Page 7
7/12
Form 1459 (01.13) ©SOR Inc.
VERIFyIng ThE CalIBRaTIOn
To verify the calibration status of a
device, begin by clicking the “Verify
calibration” button. Follow the guided
steps to complete the verification. If the
verification results are not acceptable,
start the “Calibration wizard” to either
adjust the calibration or re-range the
device.
NOTE: It may be desirable to save an
archive of the device calibration when the
device has been calibrated to a satisfactory
zero and span output. To do this, press the
“Save Report” button. You will be prompted
for a directory to save the file. Once the
directory is chosen, it will automatically be
recalled upon subsequent saves.
After the “Finish” button is pressed, the following
confirmation will be displayed to confirm the
calibration process was completed successfully.