Other features and settings – SOR Calibration Manager Software User Manual
Page 11

11/12
Form 1459 (01.13) ©SOR Inc.
Re-Zero Feature
The Manual Zero Adjust feature can be used to adjust the Zero
Set Point to compensate for Zero Set Point shifts associated
with environmental effects. For example, a pressure transmit-
ter that was calibrated at the Lenexa, KS (USA) factory may
read less than 4mA at zero pressure when placed in service at a
high-altitude location. To correct this, the output at zero can be
adjusted back to the factory calibrated zero output.
NOTE: In order for the Manual Zero Adjust feature to be enabled for use, the device
must be successfully calibrated prior to usage of this feature. If a calibration has
been started, but not successfully completed, the Manual Zero Adjust feature will
not be available until a successful calibration has been completed.
Other Features and Settings
Changing the Display units
The display units for the application can be changed to English Units or Metric Units
for convenience. To change the display units, go to View on the menu bar and select
the preferred display units.
Obtaining Software Version Information
To view application information regarding SOR Inc., application revision, etc, go to
Help on the menu bar and click “About .”
Warranty
The Calibration Manager software and
the Calibration Kit hardware are covered
by SOR warranty for six months. For
more information please contact
Customer Service.