4 brief operating instructions – Martin Christ RVC 2-25 CDplus User Manual
Page 77

RVC 2-25 CD&
11 Appendix
Version 11/2008, Rev. 2.2 of 04/12/2013 • sb
77
Translation of the original operating manual
Pos: 220 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 25 CD plus/1 10 An han g/11 0-0 040 Kurz bedi enu ngsa nleitu ng @ 8\m od_ 131 5902 622 970 _68. docx @ 50 561 @ 2 @ 1
11.4 Brief operating instructions
1. Switch on the system at the mains power switch down on the right side
of the unit.
2. Switch on the cooling trap and/or the vacuum pump.
3. Close the lid.
4. Starting the warm-up phase:
• Enter the set values for the warm-up phase into "menu" – "options" –
"Timer warm-up" und "Temperature warm-up", confirm your entries
and quit the menu.
• Press the left-hand function key "mode".
• Select the "Warm-up" menu and confirm.
• Open the lid when the warm up phase has ended.
• Install the rotors and close the lid.
5. Changing the set values for time, temperature, vacuum and safety
pressure:
• Select "menu" with the right-hand function key.
• Select "Change set values for manual mode" from the main menu.
• Select the set value by using the up/down keys and confirm with the
right-hand function key "edit".
• Change the set value by using the up/down keys and confirm with
the right-hand function key "ok".
• Quit the menu with the left-hand function key.
6. Starting the process "Evaporation manual" or "Evaporation program":
• Open the menu "Start with phase..." by pressing the left-hand
function key "mode".
• Select "Evaporation manual" or "Evaporation program" using the
up/down keys.
• Start the evaporation process by pressing the right-hand function
key.
• The function "Evaporation program" can only be selected if a
program has been created before.
The start of the evaporation process starts the following automatic
process run:
• The rotor speed increases (rotating direction left)
• At 70% of the selected maximum speed: the aeration valve closes
automatically (to avoid delays in boiling)
• At 80% of the selected maximum speed: the stop valve opens the
connection to the vacuum pump
• Further speed increase until the preselected maximum rotor speed
is reached.