11.3 Overview of evaporation times
Vessel
Number
of
samples
Sample
volume
(ml)
Temperature
(°C)
Pump
Cooling
trap
Total time
(min)
Water
10 ml test tube
24
9
45
MZ 2C
yes
635
10 ml test tube
24
9
60
MZ 2C
yes
510
1.5 ml Eppendorf caps
36
1
30
MZ 2C
yes
330
1.5 ml Eppendorf caps
36
1
45
MZ 2C
yes
270
1.5 ml Eppendorf caps
36
1
60
MZ 2C
yes
215
1.5 ml Eppendorf caps
36
1
30
KnF
no
330
1.5 ml Eppendorf caps
36
1
45
KnF
no
290
1.5 ml Eppendorf caps
36
1
60
KnF
no
230
Toluol
10 ml test tube
24
9
30
MZ 2C
yes
120
C
6
H
5
CH
3
10 ml test tube
24
9
45
MZ 2C
yes
91
10 ml test tube
24
9
60
MZ 2C
yes
65
1.5 ml Eppendorf caps
36
1
30
MZ 2C
yes
40
1.5 ml Eppendorf caps
36
1
45
MZ 2C
yes
40
1.5 ml Eppendorf caps
36
1
60
MZ 2C
yes
30
Acetonitrile
10 ml test tube
24
9
30
MZ 2C
yes
120
C
2
H
3
N
10 ml test tube
24
9
45
MZ 2C
yes
100
10 ml test tube
24
9
60
MZ 2C
yes
80
1.5 ml Eppendorf caps
36
1
30
MZ 2C
yes
45
1.5 ml Eppendorf caps
36
1
45
MZ 2C
yes
40
1.5 ml Eppendorf caps
36
1
60
MZ 2C
yes
30
1,4 Dioxane
10 ml test tube
24
9
30
MZ 2C
yes
165
C
4
H
8
O
2
10 ml test tube
24
9
45
MZ 2C
yes
110
10 ml test tube
24
9
60
MZ 2C
yes
80
1.5 ml Eppendorf caps
36
1
30
MZ 2C
yes
55
1.5 ml Eppendorf caps
36
1
45
MZ 2C
yes
45
1.5 ml Eppendorf caps
36
1
60
MZ 2C
yes
35