Martin Christ RVC 2-25 CDplus User Manual
Page 35
RVC 2-25 CD&
6 Operation
Version 11/2008, Rev. 2.2 of 04/12/2013 • sb
35
Translation of the original operating manual
correct
incorrect
Fig. 14: Permissible loading of swing-out rotors with low capacity
Pos: 126 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 127 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /06 0 Betri eb/0 60- 004 0-0 030 -00 10 Gefä ße @ 8\ mod _13 094 988 287 79_ 68.d ocx @ 468 22 @ 4 @ 1
6.4.3.1 Tubes
• Load the tubes outside of the rotational vacuum concentrator. Solvents
in the buckets or rotor blocks cause corrosion.
• Fill the tubes carefully and arrange them according to their weight.
• Follow the safety instructions and hazard warnings(see chapter 3 -
"Safety")!
Pos: 128 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 129 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /06 0 Betri eb/0 60- 004 0-0 030 -00 20 A ufna hme blöcke @ 8\ mo d_1 3094 990 493 32_ 68.d ocx @ 468 29 @ 4 @ 1
6.4.3.2 Rotor blocks
• Always load the rotor blocks symmetrically with the same accessories
and fill to avoid imbalance.
correct
incorrect
Fig. 15: Symmetrical loading of the rotor blocks
Pos: 130 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1