7 Malfunctions and error correction
52
Version 11/2008, Rev. 2.2 of 04/12/2013 • sb
Translation of the original operating manual
Error
Possible reason
Correction
Permissible imbalance exceeded
• The rotor will be stopped
immediately, the process will
also be stopped.
– improper loading of the
rotor
– glass breakage
– uneven evaporation
• Balance the load and restart the
rotational vacuum concentrator.
If the error occurs again, contact
the after-sales service (see
chapter 7.2 - "Service contact")
VSK3000 calibration error
• Incorrect calibration of the
vacuum sensor VSK3000
• The sensor must be recalibrated
(see separate operating manual)
Pos: 160 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 161 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 162 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 002 0 Pro zess meldu nge n ( auß er 2-1 8) @ 8\m od_ 130 994 030 466 0_68 .docx @ 4 721 7 @ 3 @ 1
7.1.2 Process messages
Error
Possible reason
Correction
Mains power failure
• A mains power failure will be
detected if the power supply is
interrupted under normal
conditions (run mode).
• The process will continue once
the mains power is available
again. The process times will be
reset to zero. The set control
values will be preserved.
Factory settings loaded
• All parameters are reset to the
delivery status of the system.
• The factory settings will be
loaded if the corresponding
menu command is executed in
the service menu or if an error
occurred in the parameter menu.
No rotor detected (only Triple TMM)
• Check the rotor for correct
installation.
• If the error occurs again, contact
the after-sales service (see
chapter 7.2 - "Service contact").
Nominal rotor speed not reached
• After two minutes, the rotor has
still not reached 90% of its
nominal speed.
– set speed is too high
– drive moves sluggishly
• Check the set speed.
• When the rotor is at a standstill
and the system is switched off,
check whether the drive moves
sluggishly, by hand. Contact the
after-sales service (see chapter
7.2 - "Service contact").
Safety pressure not reached
• The safety pressure inside the
rotor chamber must be reached
within 5 minutes after the start
of an evaporation process.
• Check whether the value that
was entered for the safety
pressure is not too low.
• Check the components for any
possible leaks.
Ice condenser temperature not
reached
• Only if the "Cooling trap control"
option is enabled:
The ice condenser temperature
must be
-25°C.
• If the temperature is not
reached, the heater will be
deactivated.