6.5.3.3 Special functions
Pressure increase test
NOTE
The pressure increase test can only be performed if a vacuum sensor is
installed.
Entering the set values
• Select "Special functions" from the main menu.
• Open the menu "Pressure increase test" and select "Parameter
pressure increase test".
• Enter the set values, confirm and quit the menu.
Performing the pressure increase test
• Select "Special functions" from the main menu.
• Open the menu "Pressure increase test" and select "Start pressure
increase test".
The pressure increase test will be performed. The test time progress and
the measured pressure increase are displayed in a graphical form.
Fig. 39: Graphical presentation of the pressure increase test
When the test is finished, an information window pops up and shows the
result of the pressure increase test.
Pos: 149 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 150 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /06 0 Betri eb/CDpl us/06 0- 0050 -00 30 -004 0 Pr ozess - un d Anla gen -Inf orm ation en @ 8\ mod _130 993 318 443 3_6 8.doc x @ 4 711 5 @ 4 @ 1
6.5.3.4 Process and equipment information
The "Process & equipment Information " menu informs the user about any
error, process, or system messages.
In the event of a message, a sound signal can be heard and the symbol
" " is displayed on the status bar. In addition, the process and equipment
information is displayed. If the user is in a menu, the window will not be
displayed until the user quits the main menu.
1
Message
2
Status of the
Information
3
Number of messages
pending
4
Function key for
quitting the menu
5
Function key for
acknowledging the
message
Fig. 40: Structure of the process & equipment info window