5.1 Installation site
Operate the RVC only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the RVC to thermal stress, e.g. by positioning it near
heat generators.
• Avoid direct sunlight (UV radiation).
Pos: 88 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 89 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20 E ner gieve rsor gun g-- -- --- -- --- --- -- --- -- --- -- --- --- -- --- - @ 8\m od_ 1309 335 833 862 _68. docx @ 46 613 @ 2 @ 1
5.2 Power supply
Pos: 90 / 200 Christ /37 1 RVC-BA (PROJEKTE) /RVC 2-2 5 CDpl us/05 0 Aufst ellun g un d Anschl uss/0 50- 002 0-0 010 Anschlu ssart RVC 2- 25 @ 19\ mod _13 742 166 701 26_6 8.d ocx @ 933 27 @ 3 @ 1
5.2.1 Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage
CHRIST rotational vacuum concentrators are units of safety class I.Units of
the type RVC 2-25 CD
plus
have a three-wire power cord with an IEC C13
connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back (see chapter 2.1.1 -
"Functional and operating elements"). This equipotential bonding screw
can be used to perform an earth conductor check.
Pos: 91 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 92 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20- 002 0 Sich eru nge n ba useits @ 8\ mod _13 093 3583 459 6_6 8.do cx @ 4662 9 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the RVC must be protected with 16 Amp G fuses that are to be
provided by the customer.