Troubleshooting – Labconco WaterPr BT Systems 90153 Series User Manual
Page 73

65
Problem
Possible causes
What to do
UV System only
•
The UV Timer is exhausted.
•
Replace the UV Lamp. See
Section
How to Replace the UV Lamp (UV
•
After a new UV Lamp has been
installed, reset the UV Timer. See
Section How to View or Reset the
UV System only
•
The UV Lamp is not installed
correctly or not installed at all.
•
Power OFF the system and reinstall
the UV Lamp.
•
If the red UV Lamp Alarm is still
displayed, then contact your
technical support.
This manual is related to the following products: