How to Calibrate the Tank Level (C04)
Before calibrating the tank level, the tank needs to be filled to the 100% level or TANK FULL.
1
Press the Main and “+“ Buttons together to
enter the menu. The Display will show C01.
2
Press the Main Button 3 times. The Display
will show C04. The highest Tank Level display
will be blink ing. This means that the highest
water level is ready to be calibrated.
3
Press the “+” and “-“ Buttons together to
enter the highest water level. The Display
will show the lowest Tank Level (red level)
display blinking.
NOTE: The highest and lowest tank level
must be calibrated together or else the
tank level is not calibrated.
Continued on next page