Installation – Labconco WaterPr BT Systems 90153 Series User Manual
Page 30

22
4
Measure the total volume of water (in
Liters) dispensed from the system using a
1 L graduated cylinder.
5
Press the “+” or “-“ Button to match the
volume Display to the volume measured.
6
To exit the menu, press and hold the
Main Button for 2 seconds.
To display the next menu option, press
the Main Button once.
The Product Water flowrate is now
calibrated.
This manual is related to the following products: