Pre installation, Installation requirements, Pre installation installation requirements – Labconco WaterPr BT Systems 90153 Series User Manual
Page 13

5
Pre Installation
Installation Requirements
Feedwater
Requirements
Type of Feedwater
Potable
Flowrate
≥ 10.7 GPH (40 LPH)
Minimum Pressure
≥ 7.3psi (0.5 bar)
Maximum Pressure
≤ 87.0 psi (6 bar)
Conductivity
< 2000 µS/cm
Temperature
5 to 35 °C
pH
4 - 10
Fouling Index
< 10
Iron
< 0.1 ppm as CaCO
3
Aluminum
< 0.05 ppm as CaCO
3
Manganese
< 0.05 ppm as CaCO
3
Free Chlorine
< 1 ppm
Langelier Saturation Index
< + 0.2
TOC
< 2000 ppb
Feedwater
Connection Needed
Feedwater Piping Connection
1/2 inch Male GAZ, NPTM or BSPM
Reject Flow
Requirement
Drain Capacity
7.9 GPH (30 LPH)
Environmental
Requirements
Indoor Use Only
Storage Temperature
5 °C < T < 40 °C
Operating Temperature 5 °C < T < 40 °C
Relative Humidity
Should not exceed 80% for temperatures below 31 °C
Should stay within 50% to 80% between 31 °C and
40 °C.
Altitude
< 9843 ft (3000 metres)
Installation Category
II
Pollution Degree
2