Purging air from the system – Labconco WaterPr BT Systems 90153 Series User Manual
Page 26

18
Purging Air from the System
A
•
At this time you should have installed the Filter Pack cartridge,
Barbed Fitting and the Clear Tubing. Air trapped in the Filter Pack
cartridge should now be purged from the system.
•
Verify that you have a full tank of RO water by viewing the Tank
Level display (A).
B
•
Press the Dispensing Button once to put the system into
DISPENSING mode (B).
C
•
Dispense water from the system tank to the 60% level
(approximately 3.6 L).
•
The system will go into FILLING TANK mode when the tank level is
below the 60% level (C).
•
When the tank level reaches the minimum level, the system will
finish dispensing water.