Labconco WaterPr BT Systems 90153 Series User Manual
Page 58

50
Sanitizing the System and the Tank
C
•
Open the front cover (this may require removal of the Final Filter) to
let the system go into STANDBY mode (C).
D
•
Remove the Vent Filter and install the clear elbow fitting from the
Sanitization Kit (D).
E
•
Introduce the male connector of the clear tubing firmly into the
elbow fitting (E).
F
•
Inject 200 ml (1 ml = 1 cc) of Hydrogen Peroxide solution (30%) into
the 6 Liter tank via the clear tubing (F).
•
Rinse the air vent port with 200 ml of purified water.
Continued on next page