Appendix e: quick chart – Labconco Biological Safety Cabinets 30468 Series User Manual

Page 71

background image

Original instructions

Product Service 1-800-522-7658

65


Appendix E:

Quick Chart

Model

30441_

30448_

30461_

30468_

Type

C1

C1

C1

C1

Cabinet Size (in feet)

4

4

6

6

Sash Opening (inches)

10

8

10

8

Starting Serial #

1

1403_

1403_

1403_

1403_

Nominal Avg. Downflow (FPM)

65+/-5

65+/-5

55+/-5

55+/-5

Nominal Average Inflow (FPM)

105+/-5

105+/-5

105+/-5

105+/-5

Airflow Volume (CFM)

400

323

570

463

Concurrent Balance Volume (CFM)

480

387

684

556

Recommended Duct Vacuum (WC)

.4

.4

.4

.4

Supply HEPA Data

Labconco P/N

3838401

3838401

3838403

3838403

Exhaust HEPA Data

Labconco P/N

3838501

3838501

3838503

3838503

Fluorescent/UV Lamp Data

Fluorescent Lamps (2 each)

F32T8
TL741

F32T8
TL741

F40T8

SP41

F40T8

SP41

UV Lamp

G30T8

G30T8

G30T8

G30T8


1.

The primary serial tag is on the lower outside edge of the front dress panel.
The secondary serial tag is located on the front of the electronics module on
the top right side of the cabinet. The first two digits of the serial number are
the year of production; the next two are the month. The next 5 digits are the
sequence of production, and the letter following the serial number is the
revision level of the cabinet.

2.

Each motor must be programmed by Labconco for the appropriate width
cabinet.