Labconco Biological Safety Cabinets 30468 Series User Manual
Page 57
Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
51
Original instructions
Wing Work Surfaces Removal:
Note: The work surfaces must be thoroughly decontaminated before
removing it from the cabinet.
1.
Lift the front edge of either wing work surface straight up by grasping the
knob handles. This will raise the front edge of the center work surface.
2.
Lift the front edge up, pivoting the work surface on its back edge until the
front edge of the work surface engages the catch located on the back wall
of the work area.
3.
Lift and pull both wing work surfaces out.
4.
Reinstall the wing work surfaces by engaging the tab on the back corner
of the work surface with the slot on the rear wall of the work area.
Reinstall the work surface by supporting it while lifting up on the catch to
release the work surface. Gently lower the center work surface, being sure
to engage the pins on the front corners of the work surface with the holes
in the grille, as shown in Figure 6-1.
Wing work surfaces
Center Work surface
Figure 6-1