Chapter 3: Getting Started

Product Service 1-800-522-7658

20

Original instructions

3.

The screen will now prompt you for the

password; it is:

Light
UV Light
Timer
Timer
OK



Note: Any other key sequence will return you to the Attention screen.

4.

The first screen allows you to select

whether the Axiom is connected to an
exhaust system or not. IT IS
CRITICAL THAT THE

THIS

CABINET HAS NO EXHAUST
CONNECTION

OPTION IS

SELECTED. Press

OK

to continue.



5.

The third screen allows you to set the sash

height at either 8 or 10 inches. IT IS
CRITICAL THAT THE YOU LEAVE
THE SASH HEIGHT AS IT WAS SET
AT THE FACTORY, UNLESS YOU
WISH TO RECERTIFY THE CABINET
AT ITS NEW SETTING
.

6.

The fourth screen allows you to select

whether the Axiom has a UV light or not. IT
IS CRITICAL THAT THE YOU LEAVE
THE UV LIGHT CONFIGURATION AS
IT WAS SET AT THE FACTORY
. Press

OK

to return to the first Configuration

screen.



The unit is now properly configured for operation connected to an exhaust
system. In this mode, the unit will display an exhaust alarm if the inlet relief
valve closes during operation.

NOTE: If you ever want to disconnect the Axiom from the exhaust system,
and use it in a recirculating mode, you must reconfigure it as such.