Chapter 3: Getting Started

Product Service 1-800-522-7658

14

Original instructions

Exhaust Connection Configuration

Note: The Axiom is configured at the factory to operate unconnected to an
exhaust system. When operating in this mode, the cabinet ignores the inlet
relief valve position. If you intend to connect the cabinet to an exhaust
system, YOU MUST RECONFIGURE IT FOR THIS INSTALLATION.

Note: If the Axiom is connected to an exhaust system, and you choose to
disconnect it, to exhaust its HEPA-filtered air back into the laboratory, YOU
MUST RECONFIGURE IT FOR THIS INSTALLATION. These
instructions follow.

Connecting the Axiom to an Exhaust System

a)

Mechanical configuration

1.

Using an appropriate ladder or platform, remove the exhaust

cover panel(s) on the top right side of the exhaust cover, and
install them over the hole(s) on the left side. Hand tighten the
wing nuts to secure it, as shown in figure 3-2.

2.

Connect the Axiom’s connection collar to the exhaust system,

ensuring the connection meets all appropriate codes and
regulations.

Exhaust cover installed over
exhaust outlet in left side of cover
top

Exhaust cover removed from the
right side of cover top

Figure 3-2