Labconco Biological Safety Cabinets 30468 Series User Manual

Page 52

background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

46

Original instructions

Allow the cabinet to operate until the display screen is shown.

Warm up screen-

Connected to an exhaust system

no exhaust system









Display screen

Wash hands and arms thoroughly with germicidal soap.

Wear appropriate personnel protective equipment (PPE).

Wipe-Down

Raise the sash to its full open position (approximately 22.6 inches or

574 mm). Mute the alarm by depressing the "OK/Mute" switch.

Wipe down the interior surfaces of the cabinet with 70% ethanol, or a

suitable disinfectant, and allow to dry.

Loading Materials and Equipment

Only load the materials required for the procedure. Do not overload

the cabinet.

Do not obstruct the front, side, or rear return air grilles.

Large objects should not be placed close together.

Slowly close the sash until it is in the correct operating position.

After loading the cabinet, wait two to three minutes to purge airborne

contaminants from the work area.