Chapter 6: Maintaining The Cabinet

Product Service 1-800-522-7658

50

Original instructions

Check all service valves, if so equipped, for proper operation.

Check the UV and fluorescent light hourmeters, and record their

readings in an operational log.

All weekly activities.

Semiannually or Annually

Have the cabinet re-certified by a qualified certification technician.

All monthly activities.

Service Operations

Center Work Surface Removal:
Note:
The work surface must be thoroughly decontaminated before
removing it from the cabinet.

1.

Lift either of the wing work surfaces up by grasping the knob handles at

either front corner.

2.

This will lift up the center work surface. Lift the front edge up, pivoting

the work surface on its back edge until the front edge of the work surface
engages the catch located on the back wall of the work area.

3.

Reinstall the work surface by supporting it while lifting up on the catch to

release the work surface. Gently lower the center work surface, being sure
to engage the pins on the front corners of the work surface with the holes
in the grille, as shown in Figure 6-1.