Chapter 3: Getting Started

Product Service 1-800-522-7658

16

Original instructions

3.

The screen will now prompt you for the

password; it is:

Light
UV Light
Timer
Timer
OK



Note: Any other key sequence will return you to the Attention screen.

4.

The first screen allows you to select

whether the Axiom is connected to an
exhaust system or not. IT IS
CRITICAL THAT THE

THIS

CABINET HAS AN EXHAUST
CONNECTION

OPTION IS

SELECTED. Press

OK

to continue.


5.

The second screen allows you to set

how long the blowers in the Axiom will
continue to operate in the event of an
exhaust system failure. The interval can
be programmed from 0-300 seconds.

Note: Consult with your facility’s safety officer or Labconco to help
establish how long the Axiom should continue to operate after an exhaust
system failure.


6.

The third screen allows you to set the sash

height at either 8 or 10 inches. IT IS
CRITICAL THAT THE YOU LEAVE
THE SASH HEIGHT AS IT WAS SET
AT THE FACTORY, UNLESS YOU
WISH TO RECERTIFY THE CABINET
AT ITS NEW SETTING
.