Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

42

Original instructions


Startup tone

Using the

and

buttons on the touchpad,

highlight the

Startup Tone

option-it will turn white

when selected, and then press

OK

. Using the

and

buttons on the touchpad, highlight either

Turned

on

or

Turned off

option. When turned on, an audible

beep will sound during the first 60 seconds of
blower operation to caution the user that the unit is
not yet ready for use.

Security Lock

Note: The Security Lock feature “locks” the keypad
to prevent unauthorized use of the cabinet. The
security lock is unlocked by holding the

button

for 5 seconds. After being turned on, the lock
activates whenever the sash is closed, or the blower
shuts off. Once unlocked, the lock must be
reactivated if it is desired.

From the Settings menu screen, use the

and

buttons on the touchpad to

highlight the

Security Lock

option-it will turn white when selected. Press

OK

to enter the Security Lock screen. Using the

and

buttons on the

touchpad, highlight either

turned on

or

turned off

-it will turn white when

selected, press

OK

. The screen will then return to the first Settings screen.


RS-232 Output Rate


Note: This selection will only work if the optional EN/RS-232 board is
installed.

This menu option selects the rate that the RS-232 board outputs data. Data
can be output at a rate of once per second, once every 10 seconds, once every
30 seconds, or once per minute.

From the Settings menu screen, use the

and

buttons on the touchpad to highlight the

RS-232

output

-it will turn white when selected. Press

OK

to enter the RS-232 output screen. Using the

and

buttons on the touchpad, highlight either

the data output rate preferred. It will turn white
when selected, and then press

OK

. The screen will

then return to the first Settings screen.