Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

45

Original instructions

4.

When the timer reaches 0:00:00, an audible alarm will sound, and the

timer will reset itself and repeat the countdown.

5.

Press the

OK

button to pause the timer.

6.

Press the

Menu

button to clear the interval timer and return to the

main timer menu.

Stopwatch Timer Operation

1.

When selected, the Stopwatch Timer menu is displayed on the LCD.

The timer defaults to 0:00:00.

2.

Press the

OK

button to start the timer.

3.

Press the

OK

button again to zero the timer.

4.

Press the

Menu

button to clear the stopwatch timer and return to the

main timer menu.

If An Airflow Alert Activates

The most common causes of an Airflow Alert are:

Blockage of the inlet grilles or exhaust outlet.

Removal of the work surface or grille during operation.

Resetting the Airflow Alert System

The Airflow Alert automatically resets to normal operation once the motor
speed has stabilized.

Working In the Biosafety Cabinet

Note: A more thorough review of using the BSC can be found in: Biosafety
in Microbiological and Biomedical Laboratories (BMBL )
, Published
by the Centers for Disease Control and Prevention
(

www.cdc.gov/biosafety/publications

).

Planning

Thoroughly understand procedures and equipment required before

beginning work.

Arrange for minimal disruptions, such as room traffic or entry into the

room while the cabinet is in use.

Start-up

Turn off UV light if included.

Slowly raise the sash until the bottom of the sash aligns with the sash

indicator decal located on the left side of the work area.

Turn on the fluorescent light and cabinet blower if the Smart-Start

features have not been activated.

Check the air grilles for obstructions.