Electrical requirements – Labconco Biological Safety Cabinets 30468 Series User Manual

Page 12

background image

Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Original instructions

Electrical Requirements

The biosafety cabinet models have the following electrical requirements:

Model #

Requirements

30441xx0x, 30448xx0x

115 VAC, 60 Hz, 16 Amps

30441xx1x, 30448xx1x

100 VAC, 50/60 Hz, 16 Amps

30441xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

30448xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

30461xx0x, 30468xx0x

115 VAC, 60 Hz, 16 Amps

30461xx1x, 30468xx1x

100 VAC, 50/60 Hz, 16 Amps

30461xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

30468xx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

Note: A dedicated outlet with an appropriate circuit breaker should be
located as close as possible to the right rear side of the cabinet, at a
height even with, or higher than, the top of the bench or stand. Models
rated for 100 or 115 VAC will require a 20 Amp outlet. Consult your
local electrical codes for properly rated circuit breakers. For safe
operation the dedicated outlet must provide the protective earthing
ground connection to the cabinet.

Note: On 100 and 115 VAC models, both electrical outlets are protected by a
ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the biosafety cabinet into a GFIC outlet.

Electrical outlets in the cabinet are restricted to 5 amps
maximum current.

Prises électriques dans l'armoire sont limitées à 5 courant
maximum ampères.

Do not use any detachable power cord that is not adequately
rated for the unit.

Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.

Table 2-1