Chapter 7:

Troubleshooting

Refer to the following table if the biosafety cabinet fails to operate properly.
If the suggested corrective actions do not solve the problem, contact
Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet blower and
lights won’t turn on

Unit not plugged into
outlet

Plug the biosafety cabinet into
appropriate electrical service.

Check connection to control box on top
of cabinet.

System Reset Switch
is Off

Turn on the System Reset Switch.

Circuit breaker(s)
tripped

Reset circuit breakers.

Keypad disconnected
or defective

Run sash sensor diagnostic and check
connections.

Blower won’t turn on Sash closed

Raise sash.


Keypad disconnected
or defective


Run sash sensor diagnostics and check
connections.


Blower wiring is
disconnected


Inspect blower wiring.

Blower motor is
defective

Replace blower motor.