Appendix C:

Specifications

Electrical Data

Model #

Requirements

3044xxx0x

115 VAC, 60 Hz, 16 Amps

3044xxx1x

100 VAC, 50/60 Hz, 16 Amps

3044xxx-20, 30, 40, 50, 60, 70

230 VAC, 50/60 Hz, 8 Amps

3046xxx0x

115 VAC, 60 Hz, 16 Amps

3046xxx1x

100 VAC, 50/60 Hz, 16 Amps

3046xxx-20, 30, 40, 50, 60, 70, 80 230 VAC, 50/60 Hz, 8 Amps

Motor Specifications

Cabinet Model

Electrical Requirements

All Cabinets, all
Voltages

2- 1/3 H.P. Electronically Commutated Motor (ECM)
120-240 VAC – 50/60 Hz,
Full Torque – 28 Oz.-Ft (3.56 N-M)
5 Full Load Amps @115VAC
2.6 Full Load Amps @230VAC
Automatic Thermal Protection

Environmental Conditions

Indoor use only.

Ambient temperature range: 41° to 104°F (5° to 40°C).

Maximum relative humidity: 80% for temperatures up to 88°F

(31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).

Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

Transient overvoltages according to Installation Categories II

(Overvoltage Categories per IEC 1010). Temporary voltage spikes on
the AC input line that may be as high as 1500V for 115V models and
2500V for 230V models are allowed.

Used in an environment of Pollution degrees 2 (i.e., where normally

only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664.