beautypg.com

Hanna Instruments HI 720 Series User Manual

Page 9

background image

16

17

Setup Item

Description

Valid Values

Default

TEMPERATURE
b.03

Temperature measure °C, °F

°C

unit (see note 32)

b.10

Temp. compensation

Linear,

Linear

algorithm (note 28)

NaCl

(IEC 746, table BII)

,

User: user defined table

b.11

Reference temp.

20°C, 25°C

25°C

(see note 28)

b.12

Temp. coefficient

0.00 to 20.00 %

2.00 %

(see notes 28, 29)

Temperature Compensation Table
b.31

Number of couples

2 to 10

2

b.32

Edit/view couple

1 to b.31 item value

1

b.33

User table: actual

0 to f.s.

see note 33

conductivity value for
the couple selected in b.32

b.34

User table:

-30.0 to 130.0°C

see note 33

temperature value for
the couple selected in b.32

Temperature Level Alarm (see Note 36)
b.41

Max temperature

-30.0 to 130.0°C

130.0°C

b.42

Min temperature

-30.0 to 130.0°C

-30.0°C

Reading Offset
b.50

Actual reading

measurement -10.0°C

measurement

(for ATC only,

to measurement +10°C (see note 15)

see notes 15, 20)

b.51

Temperature offset

-10.0°C to +10.0°C

0.0°C

(ATC only, notes 15, 20)

CONCENTRATION
d.00

Concentration curve

1 to 4

1

in use

Curve 1/2/3/4 Table (see note 34)
d.01

Number of triplets

1 to 25

1

d.02

Edit/view triplet

1 to d.01 value

1

Setup Item

Description

Valid Values

Default

GENERAL SETTINGS
G.00

Measurement mode

Conductivity,

Conductivity

(see note 8)

Concentration, TDS

G.01

Conductivity range

Auto range,

Auto range

1999 μS/cm,
19.99 mS/cm,
199.9 mS/cm,
2000 mS/cm

G.02

TDS range

Auto range,

Auto range

1000 ppm, 10.00 ppt,
100.0 ppt, 1000 ppt

G.03

Concentration unit

Auto (ppm/ppt),

Auto

(see note 31)

ppm, ppt, %

(ppm/ppt)

G.04

Concentration

XXXX, XXX.x,

XX.xx

format

XX.xx, X.xxx

G.05

TDS factor

0.00 to 1.00

0.50

G.06

Moving average

1 to 30

1

length (see note 30)

G.10

Factory ID

0000 to 9999

0000

G.11

Process ID alias

00 to 99

00

RS485 address

G.13

Language

English, Italiano

English

G.14

LCD contrast

0 to 10

6

G.15

LCD backlight

On, Off

On

G.98

Calibration

0000 to 9999

0000

password (see note 2)

G.99

General password

0000 to 9999

0000

(see note 2)

TEMPERATURE
b.01

Compensation mode Automatic (ATC),

ATC

(see notes 7, 20)

Manual (MTC)

b.02

Start manual temp.

-30.0 to 130.0°C

25.0°C

(see notes 7, 20)

This table lists all available setup items together with their code, short descrip-