Hanna Instruments HI 720 Series User Manual
Page 10
18
19
Setup Item
Description
Valid Values
Default
Setpoint 2
C.24
Setpoint 2 reset time
0.1 to 999.9 minutes
999.9 min.
C.25
Setpoint 2 rate time
0.0 to 999.9 minutes
0.0 min.
Alarms
C.30
Low alarm value
0 to f.s
100 μS/cm
(LA) (see note 18)
(see note 1)
C.31
High alarm value
0 to f.s
1900 μS/cm
(HA) (see note 18)
(see note 1)
C.32
Max relay ON time
1 to 60 minutes
60 min.
(then an alarm is generated)
C.33
Alarm mask time
00:00 to 30:00 min.
00:30
(see notes 18, 27)
C.34
Alarm hysteresis
0 to f.s.
20 μS/cm
(AH) (see note 18)
Programmable Control
C.41
Hold time start (note 6) 00:00 to 23:59
00:00
C.42
Hold time stop (note 6) 00:00 to 23:59
00:00
C.51
Monday hold mode
Off: disabled,
Off
On: enabled
C.52
Tuesday hold mode
Off: disabled,
Off
On: enabled
C.53
Wednesday hold
Off: disabled,
Off
mode
On: enabled
C.54
Thursday hold
Off: disabled,
Off
mode
On: enabled
C.55
Friday hold mode
Off: disabled,
Off
On: enabled
C.56
Saturday hold mode
Off: disabled,
Off
On: enabled
C.57
Sunday hold mode
Off: disabled,
Off
On: enabled
Other Parameters
C.60
Proportional control
01:00 to 30:00 min.
05:00 min.
period (if available)
Setup Item
Description
Valid Values
Default
Curve 1/2/3/4 Table (see note 34)
d.03
Not compensated
0 to f.s.
0 μS/cm
conductivity value for
the triplet selected in d.02
d.04
Temperature value for -30.0 to 130.0°C
0.0°C
the d.02 triplet
d..05
Concentration value
0 to f.s
0 ppm
for the d.02 triplet
CONTROL
C.00
Control enable
On, Off
Off
Setpoint 1
C.10
Setpoint 1 mode
Disabled,
On/Off low
(M1)
On/Off high,
(see note 1)
On/Off low,
PID high (if available),
PID low (if available)
C.11
Setpoint 1 value
0 to f.s.
500 μS/cm
(S1)
(see note 1)
C.12
Setpoint 1
0 to f.s.
20 μS/cm
hysteresis (H1)
(see note 1)
C.13
Setpoint 1
Up to f.s and
≠0
20 μS/cm
deviation (D1)
(see note 1)
C.14
Setpoint 1 reset time
0.1 to 999.9 minutes
999.9 min.
C.15
Setpoint 1 rate time
0.0 to 999.9 minutes
0.0 min.
Setpoint 2
C.20
Setpoint 2 mode
Disabled,
On/Off low
(M2)
On/Off high,
(see note 1)
On/off low,
PID high (if available),
PID low (if available)
C.21
Setpoint 2 value
0 to f.s.
1500 μS/cm
(S2)
(see note 1)
C.22
Setpoint 2
0 to f.s.
20 μS/cm
hysteresis (H2)
(see note 1)
C.23
Setpoint 2
Up to f.s and
≠0
20 μS/cm
deviation (D2)
(see note 1)