beautypg.com

Hanna Instruments HI 720 Series User Manual

Page 12

background image

22

23

Setup Item

Description

Valid Values

Default

COMMUNICATION
P.14

Remaining SMS

000 to 200, or 222

100

(see note 22)

P.15

Repeated SMS

0 to 5 (see note 23)

2

P.16

SMS delay between

05 to 60 minutes

10 minutes

messages (see note 24)

P.17

SIM card expiration

01 to 31

01

day (see note 25)

P.18

SIM card expiration

01 to 31

01

month (see note 25)

P.19

SIM card expiration

2000 to 2099

2010

year (see note 25)

CLEANING
Simple Cleaning
L.00

Rinsing time

5 to 99 seconds

20 seconds

L.01

Pause time

10 to 9999 minutes

1440 min.

Advanced Cleaning (see Note 19)
L.10

Pre-rinsing time

0 to 99 seconds

20 seconds

L.11

Cleaning time

0 to 99 seconds

10 seconds

L.12

Rinsing time

5 to 99 seconds

20 seconds

L.13

Pause time

10 to 9999 minutes

1440 min.

L.14

Minimum pause time 10 to L.13 minutes

10 minutes

L.15

Cleaning trigger

Timer only,

Timer only

External only,
Timer and external,
Timer masked by external

L.16

Repeated cycles no.

0 to 10

0

L.17

Economy cycle no.

0 to 10

0

ERROR CONFIGURATION (see Note 5)
E.00

High alarm (00)

Alarm relay -------------> On
22 mA fault current----> On
3.6 mA fault current---> Off
Auto cleaning ----------> Off
SMS sending -----------> Off

Setup Item

Description

Valid Values

Default

INPUT
I.00

Input selection

Inductive probe or

Inductive

(note 10)

Digital transmitter

probe

I.03

Digital transmitter

00 to 99

00

address (active if I.00=Digital transmitter)

I.04

Digital transmitter

1200, 2400, 4800,

19200 bps

baud rate

9600 or 19200 bps

I.10

Calibration timeout

0 to 12 months

12 months

I.11

Life check

Disabled,

Disabled

(see notes 9, 20)

1 hour, 2 hours,
or 4 hours

REAL TIME CLOCK (see Note 4)
r.00

Current day

01 to 31

from RTC

r.01

Current month

01 to 12

from RTC

r.02

Current year

2000 to 2099

from RTC

r.03

Current time

00:00 to 23:59

from RTC

COMMUNICATION
P.00

Connection type

PC,

PC

HI 504900 GSM module,
HI 504901 GSM supervisor,
HI 504902 Modem

P.01

Baud rate

1200, 2400, 4800,

19200 bps

9600 or 19200 bps

P.02

Call answer enable

On, Off

Off

P.03

Modem country

Insert the phone

000

code

country code for the
HI 504902 modem
(see note 35)

P.11

PIN code

0000 to 9999

0000

P.12

Phone number #1

+ --------------------

(see note 21)

(“Not set” means no number entered)

P.13

Phone number #2

+ --------------------

(see note 21)

(“Not set” means no number entered)