beautypg.com

Hanna Instruments HI 720 Series User Manual

Page 15

background image

28

29

Error

Alarm

22 mA

3.6 mA

Hold

Auto-

SMS

Config. Relay

fault curr. fault curr. mode

cleaning sending

22

OFF

OFF

ON

ON

ON

OFF

23

ON

OFF

ON

ON

ON

OFF

24

OFF

OFF

OFF

OFF

OFF

ON

25

ON

OFF

OFF

OFF

OFF

ON

26

OFF

ON

OFF

OFF

OFF

ON

27

ON

ON

OFF

OFF

OFF

ON

28

OFF

OFF

ON

OFF

OFF

ON

29

ON

OFF

ON

OFF

OFF

ON

30

OFF

OFF

OFF

ON

OFF

ON

31

ON

OFF

OFF

ON

OFF

ON

32

OFF

ON

OFF

ON

OFF

ON

33

ON

ON

OFF

ON

OFF

ON

34

OFF

OFF

ON

ON

OFF

ON

35

ON

OFF

ON

ON

OFF

ON

36

OFF

OFF

OFF

OFF

ON

ON

37

ON

OFF

OFF

OFF

ON

ON

38

OFF

ON

OFF

OFF

ON

ON

39

ON

ON

OFF

OFF

ON

ON

40

OFF

OFF

ON

OFF

ON

ON

41

ON

OFF

ON

OFF

ON

ON

42

OFF

OFF

OFF

ON

ON

ON

43

ON

OFF

OFF

ON

ON

ON

44

OFF

ON

OFF

ON

ON

ON

45

ON

ON

OFF

ON

ON

ON

46

OFF

OFF

ON

ON

ON

ON

47

ON

OFF

ON

ON

ON

ON

Note that the values in this table are used for error configuration through RS485.

( 6 )

The hold mode is never enabled by the control timing if the “hold time

start” is the same as the “hold time end”. Items C.41 and C.42 apply to all days.
The hold mode can be enabled all day long by using items C.51 to C.57.

( 7 )

See “Temperature Compensation” section for more details about Auto-

matic and Manual temperature compensation functioning.

( 8 )

Whenever the measurement type (G.00) is changed, all conductivity and

( 3 )

When a wrong setup code or value is confirmed, the controller does not

move from the current window, and displays a blinking WRONG message till the
user changes the value. Note that the allowed values for some parameters de-
pend from other settings (e.g. to set a high setpoint to 10.00 mS, first set the
high alarm to a value greater than 10.00 mS).

( 4 )

Each time the controller is powered, the RTC is checked to detect if an RTC

reset occurred since last software initialization. If that happened, the RTC is ini-
tialized with the default date and time (01-01-2000 ; 00:00).
An EEPROM reset does not affect the RTC settings.

( 5 )

The error configuration value is always displayed together with each error

configuration item, and is coded as shown on this table

Error

Alarm

22 mA

3.6 mA

Hold

Auto-

SMS

Config. Relay

fault curr. fault curr. mode

cleaning sending

00

OFF

OFF

OFF

OFF

OFF

OFF

01

ON

OFF

OFF

OFF

OFF

OFF

02

OFF

ON

OFF

OFF

OFF

OFF

03

ON

ON

OFF

OFF

OFF

OFF

04

OFF

OFF

ON

OFF

OFF

OFF

05

ON

OFF

ON

OFF

OFF

OFF

06

OFF

OFF

OFF

ON

OFF

OFF

07

ON

OFF

OFF

ON

OFF

OFF

08

OFF

ON

OFF

ON

OFF

OFF

09

ON

ON

OFF

ON

OFF

OFF

10

OFF

OFF

ON

ON

OFF

OFF

11

ON

OFF

ON

ON

OFF

OFF

12

OFF

OFF

OFF

OFF

ON

OFF

13

ON

OFF

OFF

OFF

ON

OFF

14

OFF

ON

OFF

OFF

ON

OFF

15

ON

ON

OFF

OFF

ON

OFF

16

OFF

OFF

ON

OFF

ON

OFF

17

ON

OFF

ON

OFF

ON

OFF

18

OFF

OFF

OFF

ON

ON

OFF

19

ON

OFF

OFF

ON

ON

OFF

20

OFF

ON

OFF

ON

ON

OFF

21

ON

ON

OFF

ON

ON

OFF