beautypg.com

A technical specifications, Fpm-600, Mportant – EXFO FPM-600 Power Meter User Manual

Page 73: Power meter/light source/optical loss test set

background image

Power Meter/Light Source/Optical Loss Test Set

67

A Technical Specifications

FPM-600

I

MPORTANT

The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at www.exfo.com
.

ORDERING INFORMATION

SPECIFICATIONS

a

Model FPM-602

FPM-602X

Detector Ge

GeX

Power range (dBm)

b

10 to –70

26 to –55

Wavelength range (nm)

800 to 1650

800 to

1650
Calibrated wavelengths (nm)

800, 820, 830, 840, 850, 860, 870, 880, 910, Same calibrated

980, 1270, 1280, 1290, 1300, 1310, 1320,

wavelengths as the

1330, 1340, 1350, 1370, 1390, 1410, 1430,

FPM-602, plus 1060

1450, 1460, 1470, 1480, 1490, 1500, 1510,

1520, 1530, 1540, 1550, 1560, 1570, 1580,

1590, 1600, 1610, 1620, 1630, 1640, 1650

Power uncertainty

c

±5 % ± 0.1 nW

±5 % ± 3 nW

Resolution (dB)

±0.01 (10 dBm to –60 dBm)

±0.01 (26 dBm to –45 dBm)

Automatic offset nulling

d

Yes

Yes

Display units

dB, dBm, W

dB, dBm, W

Tone detection

270 Hz, 1 kHz and 2 kHz

270 Hz, 1 kHz and 2 kHz

Auto-switching

e

Yes

Yes

Warm-up period (min)

f

0

0

Data storage (items)

more than 1000

more than 1000

Battery life (hours) (typical)

72

72

Warranty and recommended
recalibration interval (years)

3

3

NOTES

a. Guaranteed unless otherwise specified. All specifications valid

at 1550 nm and 23 °C ± 1 °C, with an FC connector.

b. In CW mode; sensitivity defined as 6 x rms noise level.

c. For calibrated wavelengths. Valid up to 20 dBm for

FPM-602X.

d. For power > –40 dBm for FPM-602, and > –25 dBm

for FPM-602X.

e. At 850 nm, 1300 nm, 1310 nm, 1490 nm, 1550 nm

and 1650 nm; for power > –50 dBm for FPM-602 and
> –40 dBm (typical) for FPM-602X.

f. For a variation of

) 0.06 dB at power levels * –40 dBm

for FPM-602 and

* –25 dBm for FPM-602X.

GENERAL SPECIFICATIONS

Size (H x W x D)

190 mm x 100 mm x 62 mm

(7

1

/

2

in x 4 in x 2

1

/

2

in)

Weight

0.48 kg

(1.1 lb)

Temperature

operating

–10 °C to 50 °C

(14 °F to 122 °F)

storage

–40 °C to 70 °C

(–40 °F to 158 °F)

Relative humidity

0 % to 95 % non-condensing

STANDARD ACCESSORIES

User guide, Certificate of Calibration, instrument stickers
in six languages, AC adapter/charger, lithium ion battery,
shoulder strap, carrying case, USB cable.

This manual is related to the following products: