Error codes and descriptions – EXFO FPM-600 Power Meter User Manual
Page 63
Troubleshooting
Power Meter/Light Source/Optical Loss Test Set
57
Error Codes and Descriptions
Error Codes and Descriptions
ER: error code displayed until you press a key.
WR: warning code displayed for 3 seconds, then unit returns to
normal.
Error
Code
Description
Solution
LIGH
Light detected while nulling
offsets. Nulling is not
performed.
Correctly place protective cap
on detector port, then retry.
FULL Storage memory full.
Delete data.
EMPt
Storage memory empty:
You pressed Recall but no
data was saved.
Last saved data deleted.
All data deleted.
Add data.
29/30/
34/36/
56/57
Calibration errors.
Contact EXFO.
18/52 Incompatible wavelengths or
power too low in PREF on
FOT-600.
Match source and power
meter wavelengths or increase
source power.
This manual is related to the following products: