beautypg.com

Error codes and descriptions – EXFO FPM-600 Power Meter User Manual

Page 63

background image

Troubleshooting

Power Meter/Light Source/Optical Loss Test Set

57

Error Codes and Descriptions

Error Codes and Descriptions

ER: error code displayed until you press a key.

WR: warning code displayed for 3 seconds, then unit returns to
normal.

Error

Code

Description

Solution

LIGH

Light detected while nulling

offsets. Nulling is not

performed.

Correctly place protective cap

on detector port, then retry.

FULL Storage memory full.

Delete data.

EMPt

Storage memory empty:

You pressed Recall but no

data was saved.

Last saved data deleted.

All data deleted.

Add data.

29/30/

34/36/

56/57

Calibration errors.

Contact EXFO.

18/52 Incompatible wavelengths or

power too low in PREF on

FOT-600.

Match source and power

meter wavelengths or increase

source power.