beautypg.com

Defining a list of favorite wavelengths – EXFO FPM-600 Power Meter User Manual

Page 28

background image

Measuring Power or Loss (FPM-600 and FOT-600)

22

FPM/FLS/FOT-600

Defining a List of Favorite Wavelengths

Defining a List of Favorite Wavelengths

You must put the wavelengths you want to use on a list of favorite
wavelengths (the FAV list). Only wavelengths on this list are available for
measurements. You may enter up to 40 favorite wavelengths.

Specifications are guaranteed for calibrated wavelengths only. For other
wavelengths, the unit will determine values based on the calibrated
wavelengths (3-point interpolation).

To add wavelengths to the FAV list (or to remove them):

1. From normal operating mode, press

repeatedly until you reach PM.

2. Press

to access the first submenu, then

press

repeatedly

until

you

reach

the

FAV

menu.

3. Press

to switch between available wavelengths. An asterisk (

*)

appears beside wavelengths already on the list. If no wavelength is set
for a position, the unit indicates “----”. You may also press

to

switch between defined wavelengths only.

Detector

Type

Calibrated

Wavelengths (nm)

Default Favorite

Wavelengths

a

(nm)

a.

To revert to default favorite wavelengths see Reverting Unit to Factory Settings on

page 56.

Ge

800, 820, 830, 840, 850, 860, 870, 880,

910, 980, 1270, 1280, 1290, 1300,

1310, 1320, 1330, 1340, 1350, 1370,

1390, 1410, 1430, 1450, 1460, 1470,

1480, 1490, 1500, 1510, 1520, 1530,

1540, 1550, 1560, 1570, 1580, 1590,

1600, 1610, 1620, 1630, 1640, 1650.

GeX

All the above plus 1060.

Same as above