Measuring Power or Loss (FPM-600 and FOT-600)
22
FPM/FLS/FOT-600
Defining a List of Favorite Wavelengths
Defining a List of Favorite Wavelengths
You must put the wavelengths you want to use on a list of favorite
wavelengths (the FAV list). Only wavelengths on this list are available for
measurements. You may enter up to 40 favorite wavelengths.
Specifications are guaranteed for calibrated wavelengths only. For other
wavelengths, the unit will determine values based on the calibrated
wavelengths (3-point interpolation).
To add wavelengths to the FAV list (or to remove them):
1. From normal operating mode, press
repeatedly until you reach PM.
2. Press
to access the first submenu, then
press
repeatedly
until
you
reach
the
FAV
menu.
3. Press
to switch between available wavelengths. An asterisk (
*)
appears beside wavelengths already on the list. If no wavelength is set
for a position, the unit indicates “----”. You may also press
to
switch between defined wavelengths only.
Detector
Type
Calibrated
Wavelengths (nm)
Default Favorite
Wavelengths
a
(nm)
a.
To revert to default favorite wavelengths see Reverting Unit to Factory Settings on
Ge
800, 820, 830, 840, 850, 860, 870, 880,
910, 980, 1270, 1280, 1290, 1300,
1310, 1320, 1330, 1340, 1350, 1370,
1390, 1410, 1430, 1450, 1460, 1470,
1480, 1490, 1500, 1510, 1520, 1530,
1540, 1550, 1560, 1570, 1580, 1590,
1600, 1610, 1620, 1630, 1640, 1650.
GeX
All the above plus 1060.
Same as above