beautypg.com

8 troubleshooting, Solving common problems, 8troubleshooting – EXFO FPM-600 Power Meter User Manual

Page 61

background image

Power Meter/Light Source/Optical Loss Test Set

55

8

Troubleshooting

Solving Common Problems

Problem

Possible Cause

Solution

The unit does not turn on.

You did not press

long enough.

AC adapter/charger

not connected.

Main battery

discharged.

Weather too cold.

Press

for

2

seconds.
Connect AC

adapter/charger and

charge battery.

The unit takes very long to

turn on.

Too many values saved

in memory.

Delete all data from

the unit.

Battery does not charge as

expected.

Temperature too

high.

Battery incorrectly

connected.

Incorrect charger

used.

Ensure temperature is

within specifications.
Ensure battery is

connected properly.
Use the correct

charger.

Unable to change power

meter wavelength.

Unit receiving Auto

(or REF) signal

from source.

Only one

wavelength in list.

Change source mode

(see Modulating the
Source Signal
on

page 34), then retry.

Add wavelengths.

This manual is related to the following products: