beautypg.com

Arning – EXFO FPM-600 Power Meter User Manual

Page 52

background image

Maintenance

46

FPM/FLS/FOT-600

Cleaning EUI Connectors

To clean EUI connectors:
1.
Remove the EUI from the instrument to expose the connector

baseplate and ferrule.

2. Moisten a 2.5 mm cleaning tip with one drop of isopropyl alcohol

(alcohol may leave traces if used abundantly).

3. Slowly insert the cleaning tip into the EUI adapter until it comes out on

the other side (a slow clockwise rotating movement may help).

4. Gently turn the cleaning tip one full turn, then continue to turn as you

withdraw it.

5. Repeat steps 3 to 4 with a dry cleaning tip.

Note: Make sure you don’t touch the soft end of the cleaning tip.

W

ARNING

Looking into the optical connector while the light source is active
WILL result in permanent eye damage. EXFO strongly recommends
to TURN OFF the unit before proceeding with the cleaning
procedure.

Push

Turn

Pull

3

4

5

This manual is related to the following products: