Troubleshooting

56

FPM/FLS/FOT-600

Reverting Unit to Factory Settings

Reverting Unit to Factory Settings

You can revert most parameters on your unit to their factory state. When
you perform this operation, you will lose all customized parameters.

To revert unit to factory settings:
1.
Turn off your unit.

2. While holding down

, press

. When your unit beeps, release

.

Unable to change power

meter dB unit or reference

power.
OR
Changed unit or reference

value are replaced by other

values after a while.

Unit receiving REF

signal from source. See
Sending Source Power
Value with Signal
on

page 36.

Wait a few seconds

until power value is

received, then retry.

Many beeps, unstable optical

power and blinking

Auto (or

modulation) indicator.

Power too low to

recognize Auto mode

(or modulation).

Increase source

power or switch

source to CW.

Reference power different

than source output power.

Received power

outside detector’s

range.

Change source output

power.

Unit displays FULL even after

you deleted a few values.

You must delete all

values from the unit to

free memory.

Delete all values as

explained in Saving,
Recalling and
Deleting Data
on

page 40.

Going from the first value to

the last value in recalled data

is very slow.

Too many values saved

in memory.

Delete all data from

the unit.

Problem

Possible Cause

Solution