Measuring Power or Loss (FPM-600 and FOT-600)

20

FPM/FLS/FOT-600

Setting Power Correction Factor

Setting Power Correction Factor

You may apply a correction factor (CF) to measured power to compensate
for inaccuracies or drifts. You should change the CF after performing an
offset nulling.

For each favorite wavelength, the CF is set to 1.00 at the factory (even if the
unit indicates “----”), but allowed values range between 0.85 and 1.15.

Note: Some other products express the CF in dB, so the CF would be added to

measured power.

To set a correction factor for one or more wavelengths:
1.
From normal operating mode, press

repeatedly until you reach PM.

2. Press

to

access

the

first

submenu,

then

press

repeatedly until you reach the CF

menu.

3. Press

to switch between available

wavelengths. An asterisk (*) appears beside activated correction
factors. If no correction factor is set for a wavelength, the unit indicates
----”.

4. Press

to activate/deactivate the displayed correction factor. A

“CF” indicator will appear next to measured power in normal mode.

Power

corrected

Power

measured

CF

=