Recalibrating the Unit

Recalibrating the Unit

Manufacturing and service center calibrations are based on the
ISO/IEC 17025 Standard, which states that calibration documents must not
contain a recommended calibration interval, unless this has been
previously agreed upon with the customer.

Validity of specifications depends on operating conditions. For example,
the calibration validity period can be longer or shorter depending on the
intensity of use, environmental conditions and unit maintenance. You
should determine the adequate calibration interval for your unit according
to your accuracy requirements.

Under normal use, EXFO recommends calibrating your unit every three
years.

Note: The FlexCare warranty program includes Calibration/Verification packages

(see Service and Repairs on page 64).

To view the last calibration date (FOT-600 and FPM-600 only):
1.
From normal operating mode, hold down

and

for a few seconds. The unit displays

the first embedded software version.

2. Press

until you reach the calibration date

(and version) of the power meter.

3. Press to return to normal mode.