Troubleshooting

44

FLS-5834A

CD/PMD Analyzer Source Error Messages

-40

FIFO not ready for reading:
The unit’s FIFO is not ready for

reading. Commands sent will be

ignored.

If problem persists, call EXFO for

assistance.

-60

Laser over-current:
A problem occurred with the current

going to the laser.

Call EXFO for assistance.

-64

EEPROM error:
The EEPROM was not detected.

Call EXFO for assistance.

-65

EEPROM checksum error:
A checksum error was detected by

the unit.

Call EXFO for assistance.

-68

Temperature error:
The operating temperature of the unit

is from 10 °C to 40 °C (50 °F to 104 °F).

Make sure the ambient

temperature remains within the

specified temperature range.

-69

FPGA problem:
An FPGA problem was detected.

Call EXFO for assistance.

-73

Supply voltage error:
Internal voltage values are out of the

operation range.

Call EXFO for assistance.

-74

Laser pump failure:
A problem occurred in the output

(optical) power of the laser pump.

Call EXFO for assistance.

-75

Laser temperature problem:
Laser temperature is out of the

operation range.

Make sure the ambient

temperature remains within the

specified temperature range.

Warning/Error

Number

Description

Recommended Action