CD/PMD Analyzer Source Error Messages

Warning/Error

Number

Description

Recommended Action

-11

Module reset error:
The nulling was not performed

correctly.

Restart your unit to solve the

problem.

-12

Wrong module ID:
The module returns the wrong ID.

Call EXFO for assistance.

-20

Module communication error:
Communication error with the

module.

Restart your unit to solve the

problem.

-25

Checksum error:
Checksum error while reading the

module’s FIFO.

Restart your unit to solve the

problem.

-30

Command not accepted:
The command that caused the

warning will be lost. The unit may

continue with the program even if the

command was not performed.

If problem persists, call EXFO for

assistance.

-31

Module setting error:
One of the settings sent to the unit is

wrong.

Review your command before

sending it again.

-32

Action currently in progress:
You cannot send a command while

the unit is already active.

Wait until the unit is done before

sending the command.

-34

Command overflow:
Too many commands were sent to

the unit at a time.

Wait until the unit is done before

sending more commands.