beautypg.com

Communication parameters – EXFO FLS-5800 CD/PMD Analyzer Source User Manual

Page 33

background image

Controlling the Source Remotely

CD/PMD Analyzer Source

27

Communication Parameters

Communication Parameters

The communication parameters are used to set the communication port.

Note: EOS means “End of String.” EOI means “End or Identify.”

For GPIB Communication

Terminate Read on EOS

Yes

Set EOI with EOS on Writes

Yes

Type of compare on EOS

8 bits

EOS byte

0Ah

Send EOI at end of Writes

Yes

GPIB primary address

See Setting the GPIB Address on

page 24

GPIB secondary address

None

For RS-232 Communication

EOS bytes

0Ah

Baud rate

1200/2400/4800/9600/ 19200 bps

Parity

None

Data bits

8 bits

Stop bits

1 bit

Flow control

Software (Xon/Xoff) or None

Activation

See Setting the Remote Command
Mode