4.5.2.3 CLT Command
The CLT command is used for the temperature calibration. With this command, the contents of CE
register 71 are read and entered into MPU register C0 and the contents are saved to flash. The CLT
command example is given below::
>CLT
The response is:
TCal OK
>
A summary of the atomic calibration commands are given in the table below:
CLxx
Atomic Calibration Commands
Description:
Allows the user to Calibrate individual sections of the IC.
Usage:
CLV
Calibrates voltage only.
CLI1
Calibrate current on Outlet 1 only.
CLI2
Calibrate current on Outlet 2 only.
CLI3
Calibrate for current on both Outlet 1 and
Outlet 2 only.
CLW1
Calibrate for power on Outlet 1 only.
CLW2
Calibrate for power on Outlet 2 only.
CLW3
Calibrate for power on Outlet 1 and Outlet 2
only.
CLT
Calibrate temperature only.
The commands that follow are mainly for advanced users and are included for reference
only. These commands are not needed to operate the 78M6612 OMU or AC-PMON demo
kits.