6612_OMU_S2_URT_V1_13 Firmware Description Document

FDD_6612_010

12

Rev. 1.1

4.5.2.3 CLT Command

The CLT command is used for the temperature calibration. With this command, the contents of CE
register 71 are read and entered into MPU register C0 and the contents are saved to flash. The CLT
command example is given below::

>CLT

The response is:

TCal OK
>

A summary of the atomic calibration commands are given in the table below:

CLxx

Atomic Calibration Commands

Description:

Allows the user to Calibrate individual sections of the IC.

Usage:

CLV

Calibrates voltage only.

CLI1

Calibrate current on Outlet 1 only.

CLI2

Calibrate current on Outlet 2 only.

CLI3

Calibrate for current on both Outlet 1 and
Outlet 2 only.

CLW1

Calibrate for power on Outlet 1 only.

CLW2

Calibrate for power on Outlet 2 only.

CLW3

Calibrate for power on Outlet 1 and Outlet 2
only.

CLT

Calibrate temperature only.

The commands that follow are mainly for advanced users and are included for reference
only. These commands are not needed to operate the 78M6612 OMU or AC-PMON demo
kits.